Tubulin
Basics
Back to protocols
I.
Useful Values
1 mg/ml tubulin = 10 µM (assuming MW of ab-tubulin heterodimer
is 100,000; in reality it is ~110,000 but almost all tubulin labs use
this convenient conversion relationship)
1 µm of a microtubule has ~1600 tubulin heterodimers
Extinction coefficient of tubulin at 280 nm = 115,000 M-1cm-1
or 1.15 (mg/ml)-1 cm-1; assuming that the MW of
a tubulin heterodimer is 100,000 daltons. This extinction coefficient
was calculated from tubulin sequences and includes the contribution
of the two bound guanine nucleotides to the absorbance at 280 nm.
II. Handling Tubulin
Tubulin is a labile
molecule that converts to a non-polymerizing state within hours on ice.
This lability demands a certain amount of care and discipline when handling
tubulin. The following is what we recommend for general use of tubulin:
1. Tubulin is very stable at -80¡C and is best stored in
small aliquots at this temperature. We generally prepare
tubulin in large scale and store large aliquots at -80¡C. We
recycle 3-4 of these large aliquots and store small aliquots of the
recycled tubulin at -80¡C for daily use. One or
more recycled aliquots is thawed for an experiment, and any leftover
material discarded. Refreezing thawed aliquots of recycled tubulin is
not recommended.
2. To use recycled tubulin, rapidly thaw the aliquot by placing
it in a 37¡C bath straight from the -80¡C freezer (or from
a liquid nitrogen dewar -- we often use liquid nitrogen to bring tubulin
aliquots from the freezer to the 37¡C bath/benchtop). When the
tubulin is nearly thawed (i.e. when there is a very small amount of
ice left in the tube) transfer the tube to an ice bucket; mix it by
gentle flicking ~20s later and store on ice. Use the thawed tubulin
for polymerization preferably within 30' of thawing. When quantitative
analysis of tubulin polymerization is necessary, careful and consistent
handling of tubulin aliquots is essential for reproducible results.
Back
to top
Back to protocols